ENA450 (PQA)

Power quality analyzer ENA450 is based on a modular National Instruments Compact RIO hardware with a configurable number of voltage and current inputs. The ENA450 mounts directly on a 35mm DIN rail.

ENA450 is available in four variants: ENA450.EB, ENA450.EC, ENA450.NB, ENA450.NC.

Key features:

  • 4 voltage and 4 direct current galvanic isolated inputs with antialiasing filters. (ENA450.EB, ENA450.EC)
  • 3 voltage and 3 indirect current (with current clamps) galvanic isolated inputs with antialiasing filters. (ENA450.NB, ENA450.NC)
  • Adaptive sampling frequency
  • A/D converter with 24-bit resolution for high precision (only for ENA450.NB, ENA450.NC)
  • Optional digital inputs/outputs
  • Consumes less than 10 VA
  • No rotating parts for high reliability
  • Operating system VxWorks

ENA450 analyzer fully complies with the IEC 61000-4-30 standard, class A.

Specification

Technical Specification ENA450.EB:

General
Storage intervals k*1 min/ 10 min/ 15 min/ 2 h
Operating temperature -20 up to +55 °C
Storage temperature -40 up to +85 °C
Size 95 x 285 x 90 mm (h x w x d)
Weight 2.1 kg
Voltage Inputs
Number of inputs 4
Input ranges 57.7 V or 100 V RMS (measurable up to 2xUn)
Bandwidth 45 Hz up to 3 kHz
Isolation 4,2 kV (50 Hz / 5 s), Ch-Ch, Ch-GND
Accuracy +/-0.1 %
Connection 1-phase, 3-phase, 3-phase+N, star, delta and Aron
Current Inputs
Number of inputs 4
Input ranges 1 A or 5 A RMS (measurable up to 2xIn)
Bandwidth 45 Hz up to 3 kHz
Isolation 4,2 kV (50 Hz / 5 s), Ch-Ch, Ch-GND
Accuracy +/-0,1 %
Digital Inputs/Outputs
Number Optional
Standards
Safety IEC 60950
EMC IEC 61000-6-2, IEC 61000-4-2 to 6
Analysis procedure EN 50160, IEC 61000-4-7, IEC 61000-4-15, IEC 61000-4-30
Hardware
Memory 128 MB RAM
Processor PowerPC 400 MHz
Storage 256 MB internal or optional 2 GB
Operating system VxWorks
Interface RS232, 2x Ethernet
A/D converter 16 bits
Sampling frequency 9600S/s adaptive
Power supply 19 up to 30 V DC, optionally from measured signals 100 V and battery backup for 5 min
Power consumption Max 10 VA

 

Technical Specification ENA450.EC:

General
Storage intervals k*1 min/ 10 min/ 15 min/ 2 h
Operating temperature -40 up to +70 °C
Storage temperature -40 up to +85 °C
Size 95 x 275 x 90 mm (h x w x d)
Weight 2.5 kg
Voltage Inputs
Number of inputs 4
Input ranges 57.7 V or 100 V RMS (measurable up to 2xUn)
Bandwidth 45 Hz up to 3 kHz
Isolation 4,2 kV (50 Hz / 5 s), Ch-Ch, Ch-GND
Accuracy +/-0.1 %
Connection 1-phase, 3-phase, 3-phase+N, star, delta and Aron
Current inputs
Number of inputs 4
Input ranges 1 A or 5 A RMS (measurable up to 2xIn)
Bandwidth 45 Hz up to 3 kHz
Isolation 4,2 kV (50 Hz / 5 s), Ch-Ch, Ch-GND
Accuracy +/-0,1%
Digital Inputs/Outputs
Number Optional
Standards
Safety IEC 60950
EMC IEC 61000-6-2, IEC 61000-4-2 to 6
Analysis procedure EN 50160, IEC 61000-4-7, IEC 61000-4-15, IEC 61000-4-30
Hardware
Memory 128 MB RAM
Processor PowerPC 400 MHz
Storage 2 GB internal
Operating system VxWorks
Interface USB, RS232, Ethernet
A/D converter 16 bits
Sampling frequency 9600S/s adaptive
Power supply 9 up to 35 V DC, optionally from measured signals 100 V and battery backup for 5 min
Power consumption Max 10 VA

 

Technical Specification ENA450.NB:

General
Storage intervals k*1 min/ 10 min/ 15 min/ 2 h
Operating temperature -20 up to +55 °C
Storage temperature -40 up to +85 °C
Size 95 x 285 x 90 mm (h x w x d)
Weight 1.8 kg
Voltage Inputs
Number of inputs 3
Input ranges 120 V, 240 V RMS (max 300 V RMS)
Bandwidth 45 Hz up to 3 kHz
Isolation 2,3 kV (50 Hz / 5 s), Ch-Ch, Ch-GND
Accuracy +/-0.1 %
Connection 1-phase, 3-phase, star, delta and Aron
Current Inputs
Number of inputs 3
Input ranges According to used current clamps
Bandwidth 45 Hz up to 3 kHz
Accuracy +/-0,1 %
Digital Inputs/Outputs
Number Optional
Standards
Safety IEC 60950
EMC IEC 61000-6-2, IEC 61000-4-2 to 6
Analysis procedure EN 50160, IEC 61000-4-7, IEC 61000-4-15, IEC 61000-4-30
Hardware
Memory 128 MB RAM
Processor PowerPC 400 MHz
Storage 256 MB internal or optional 2 GB
Operating system VxWorks
Interface RS232, 2x Ethernet
A/D converter 24 bits
Sampling frequency 9600S/s adaptive
Power supply 19 up to 30 V DC, optionally from measured signals 100 V and battery backup for 5 min
Power consumption Max 10 VA

 

Technical Specification ENA450.NC:

General
Storage intervals k*1 min / 10 min / 15 min / 2 h
Operating temperature -40 up to +70 °C
Storage temperature -40 up to +85 °C
Size 95 x 180 x 90 mm (h x w x d)
Weight 1.5 kg
Voltage Inputs
Number of inputs 3
Input ranges 120 V or 240 V RMS (max 300 V RMS)
Bandwidth 45 Hz up to 3 kHz
Isolation 2,3 kV (50 Hz / 5 s), Ch-Ch, Ch-GND
Accuracy +/-0.1 %
Connection 1-phase, 3-phase, star, delta and Aron
Current Inputs
Number of inputs 3
Input ranges According to used current clamps
Bandwidth 45 Hz up to 3 kHz
Accuracy +/-0,1 %
Digital Inputs/Outputs
Number Optional
Standards
Safety IEC 60950
EMC IEC 61000-6-2, IEC 61000-4-2 to 6
Analysis procedure EN 50160, IEC 61000-4-7, IEC 61000-4-15, IEC 61000-4-30
Hardware
Memory 128 MB RAM
Processor PowerPC 400 MHz
Storage 2 GB internal
Operating system VxWorks
Interface USB, RS232, Ethernet
A/D converter 24 bits
Sampling frequency 9600S/s adaptive
Power supply 9 up to 35 V DC, optionally from measured signals 100 V and battery backup for 5 min
Power consumption Max 10 VA